Black Friday Early Deals Now Live!
VIEW DEALS

Syringe Pump Driver KL602



Syringe Pump Driver KL602
The Syringe Pump Driver KL602 is a time-based infusion pump that is designed to infuse drugs into patients. It has a flow rate of 10 ml/h, 20 ml/h, 30 ml/h, 50 ml/h, and 60 ml/h, and a bolus rate of 400 ml/h. It has an accuracy of 2% (mechanical accuracy 1%), and a body weight limit of mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc. The Syringe Pump Driver KL602 is compatible with any standard syring... more details
Key Features:
  • Time-based infusion pump
  • 10 ml/h, 20 ml/h, 30 ml/h, 50 ml/h, 60 ml/h flow rates
  • 400 ml/h bolus rate


R12 255.90 from Cleansafe Labs

price history Price history

   BP = Best Price   HP = Highest Price

Current Price: R12 255.90

loading...
Features
Brand Unbranded
Manufacturer Unbranded
Model Number KL602
Description
The Syringe Pump Driver KL602 is a time-based infusion pump that is designed to infuse drugs into patients. It has a flow rate of 10 ml/h, 20 ml/h, 30 ml/h, 50 ml/h, and 60 ml/h, and a bolus rate of 400 ml/h. It has an accuracy of 2% (mechanical accuracy 1%), and a body weight limit of mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc. The Syringe Pump Driver KL602 is compatible with any standard syringe, and has an infusion mode that allows for flow rate of ml/min, ml/h. It also has a time-based infusion mode, and an alarm that sounds when the system detects an occlusion, near-empty state, or when the battery is low. The Syringe Pump Driver KL602 also has an additional feature that real-time displays the infused volume, and has an automatic power switch that turns the pump on when it is plugged in, and off when it is not in use. The Syringe Pump Driver KL602 is also compatible with the Drug Library, and has an occlusion sensitivity

Syringe Pump KL602
Specifications
Model KL-602
Syringe Size 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 1000 ml in 1 ml increments
Flow Rate Syringe 10 ml: 0.1-400 ml/h
Syringe 20 ml: 0.1-600 ml/h 100 ml/h in 1 ml/h increments Syringe 50/60 ml: 0.1-1300 ml/h
Bolus Rate Syringe 10 ml: 400 ml/h Syringe 20 ml: 600 ml/h Syringe 30 ml: 900 ml/h Syringe 50/60 ml: 1300 ml/h
Anti-Bolus Automatic
Accuracy 2% (mechanical accuracy 1%)
Infusion Mode Flow rate: ml/min, ml/h
Time-based
Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc.
KVO Rate 0.1-1 ml/h (in 0.1 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm,
pressure sensor error, syringe installation error, syringe drop off
Additional Features Real-time infused volume, automatic power switching,
automatic syringe identification, mute button, purge, bolus, anti-bolus,
system memory
Drug Library Available
Occlusion Sensitivity High, medium, low
Power Supply, AC 110/230 V (optional), 50/60 Hz, 20 VA
Battery 9.61.6 V, rechargeable
Battery Life 7 hours at 5 ml/h
Working Temperature 5-40
Relative Humidity 20-90%
Atmospheric Pressure 860-1060 hpa
Size 314*167*140 mm
Weight 2.5 kg
Safety Classification Class II, type CF
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.