Description
The Aed Saver 1 is a semi-automatic defibrillator that is available in both a fully automatic and a fully automatic with LCD carrycase version. The Aed Saver 1 is capable of delivering a standard maximum of 200J or a power maximum of 360J of energy. The Aed Saver 1 uses an adaptive BTE waveform that conforms to the patient's chest impedance. There are various adult shock protocols available on request, and the factory default is an escalating 150, 200, 200J shock. The Aed Saver 1 has a charging time of 9 seconds and an analysis time of 4 to 15 seconds. The Aed Saver 1 has a sensitivity of IEC/EN 60601-2-4 (AHADB, MITDB source), 97% and a specificity of IEC/EN 60601-2-4 (AHADB, MITDB source), 99%. The Aed Saver 1 has two buttons for automatic operation: ON/OFF and an i info button. There are three buttons for semi-automatic operation: ON/OFF, shock button, and i info button. The Aed Saver 1 has a flashing Icons indicator, a status LED indicator, and an audible alert and text
DEFIBRILLATOR
Operation: Semi- Automatic Version
Fully Automatic Version
Energies: Standard max 200J or Power max 360J
Waveform: Adaptive BTE (biphasic truncated exponential) conforming to patient chests impedance
Protocols: Various adult shock protocols available on request
Factory default: Adult Standard escalating 150, 200, 200J
Adult Power escalating 200, 250, 360J
Paediatric (Standard or Power) 50J fixed
Charging time: 9 seconds with a new and fully charged battery depleted battery will result in a longer charging time
Analysis time: IEC/EN 60601-2-4 from 4 to 15 seconds
Impedance: 20-200 ohms
Sensitivity: IEC/EN 60601-2-4 (AHADB, MITDB source), 97%
Specificity: IEC/EN 60601-2-4 (AHADB, MITDB source), 99%
Controls: 2 buttons for Automatic: ON/OFF, i info button
3 buttons for Semi-Automatic: ON/OFF,
i info button, shock button
Flashing Icons: connect pads to patient
adult/child informing on pads type use
dont touch patient warning to stay clear
touch patient informing its safe to touch
Indicators: Status LED indicator informing on device condition
Battery gauge with remaining capacity rate
Audible alerts and text display with service alarms
Upgradeable: through a USB cable or memory card
PADS OPTIONS
Type: Disposable, pre-gelled and self-adhesive
Adult: Code SAV-C0846, for patient >8 years or >25 kg
Paediatric: Code SAV-C0016, for patient