Description
The patient monitor CMS8000 veterinary is a device that is designed to monitor standard parameters such as ECG, RESP, SpO2, PR, NIBP, and TEMP. It also has a variety of other functions such as arrhythmia and ST-segment analysis, drug concentration calculation, and alarm event marking. It is a fanless design that is quiet and energy-saving, and it has a multi-language interface. It has a standard interface that includes OxyCRG, trend chard, and a big character interface. It is operated by keys and buttons. The device has a display that is capable of displaying 8-channel waveform on one screen. It also has a display that is capable of displaying 7-lead ECG waveform on one screen. It has anti-motion and anti-ambient light interference and can measure under the circumstance of weak filling. The NIBP measurement mode allows for storage for 4800-group NIBP data. The device has a review for 71 alarm events of all parameters and 60 arrhythmia alarm events. It also has drug concentration calculation and titration table functions. The device can connect to a central monitoring system by 3G, WiFi, or wired mode. It
Patient Monitor CMS8000 Veterinary
Introduction
The CMS8000VET Veterinary Monitor has abundant monitoring functions and can be used for monitoring of cat, dog and other animals. User may select different parameter configurations according to different requirements. It adopts 100-240V,50/60Hz for power supply, 12.1'' TFT color LCD for displaying real-time data and waveform, and up to 8-channel waveform and all monitoring parameters can be displayed simultaneously. In addition, it can be connected to the central monitoring system via wired/wireless network so as to form a network monitoring system.
This device can monitor such parameters as ECG, RESP, NIBP, SpO2 and dual- channel TEMP, etc. It integrates parameter measurement module, display and recorder in one device to form a compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for moving.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate (HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc
Features
1) 12.1'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).
2) Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
3) With standard interface, OxyCRG, trend chard, big character interface and view bed, convenient to observe.
4) Operation mode: keys and buttons.
5) Display up to 8-channel waveform on one screen.
6) Display 7-lead ECG waveform on one screen, and with ECG step function.
7) Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
8) NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
9) Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
10) Drug concentration calculation and titration table functions.
11) Print trend table data with one-key.
12) Connect to Central Monitoring System by 3G, WiFi or wired mode.
13) AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
14) Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Performance
ECG
1.Lead mode:3-lead or 5-lead
2.Lead select:I, II, III, aVR, aVL, aVF, V
3.Waveform: 5-lead, 2-channel/3-lead, 1-channel
4.Gain:2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
5.Scan speed:12.5mm/s, 25 mm/s, 50 mm/s
HR:
1.Measurement and alarm range: 15~350bpm
2.Accuracy:1% or 1bpm, whichever is greater
3.Alarm accuracy: 2 bpm
4.Resolution:1 bpm
5.ST-segment monitoring:
6.Measurement and alarm range: -2.0mV~+2.0mV
7.Accuracy:-0.8mv~+0.8mv 0.04mv or 10%, whichever is greater
8.Other range: unspecified
9.Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
10.Pacemaker:yes
RESPIRATION
1.Method:R-F(RA-LL) Impedance
2.Respiration rate:
Measurement and alarm range: 0~150rpm
3.Resolution:1 rpm
4.Measurement accuracy: 2 rpm
5.Alarm accuracy:3 rpm
6.Apnea alarm:10~40s
7.Scan speed:6.25 mm/s, 12.5 mm/s, 25 mm/s
NIBP
1.Method:Oscillometry
2.Mode:Manual/AUTO/STAT
3.Measurement interval in AUTO mode: 4.1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
5.Measurement period in STAT mode: 5 minutes
6.Measurement and alarm range:10 ~ 270 mmHg
7.Resolution:1 mmHg
8.Cuff pressure accuracy:3 mmHg
9.Measurement accuracy:
10.Maximal mean error: 5 mmHg
11.Maximal standard deviation: 8mmHg
12.Over-pressure protection:
13.Other mode:2973 mmHg
Dog mode:2403 mmHg
Cat mode:1473 mmHg
SpO2
1.Measurement and alarm ra